We are offering Brass Body Oxygen Terminal In Various Standards. These products are available with us in various standards and specifications. These products are made by using high quality brass material.This is a BRASS BODY OXYGEN TANK TERMINAL, which can be connected to wall-type oxygen regulator for respiratory supply. Mostly used in hospital and nursing home. This is made of brass material and customized according to customor requirement.
The brass body oxygen terminal is used to connect to wall-style oxygen regulator for respiratory supply. It is made of high quality brass and have the ability of providing consistency, reliability and safety for your needs.Oxygen Terminal is made of brass material, which looks like a valve, and it has the function of controlling the flow of oxygen. When the oxygen passes through this component, the high pressure is lowered down to a safe level before entering into human respiratory system. This product can be customized in nearly all sizes and styles as per requirement of customers.
Oxygen is a gas that is essential for sustaining life in human beings. Oxygen is delivered to the body through a tube connected to an oxygen cylinder. Usually, you will find an oxygen retention when you need this gas. The most common types of these retention include mechanical, hydraulic and pneumatic tubes. The best way of dealing with this retention is through the use of medical tubing that can be easily obtained at pharmacies or medical supplies stores.
Introduction to Terminal Socket Series
1. Intended use
This series of terminal sockets (hereinafter referred to as terminals) are supporting devices used for gas supply in hospital wards and operating rooms in medical center gas supply systems. The flow media include oxygen, air, laughing gas, nitrogen, carbon dioxide, etc.
2. Technical parameters
2.1 Applicable ambient temperature: -20 ºC~40 ºC, ambient humidity: 10%~95%;
2.2 Working pressure: -0.1~0.8MPa;
2.3 The maximum gas supply flow rate is greater than 100L/min; The maximum inspiratory flow rate is greater than 40L/min.
2.4 Executive standard: Q/CTL003-2006
3. Precautions for installation and use
Before installing and using this terminal, check if the terminal is intact. The defect of the parts will cause harm to users and others, and it is strictly prohibited to use the terminal parts when they are damaged.
When installing and using this terminal, the gas used can only be medical gas, and the pressure cannot exceed the working pressure.
The installation of this terminal does not allow the use of grease like items. If cleaning and leak detection are required, household neutral detergents can be used for leak testing or cleaning. Do not use any organic solvents, synthetic fats, chemical reagents or other oily or non neutral liquids for leak testing or cleaning. These fatty substances and oxygen can cause combustion or accelerate intense combustion or explosion.
When installing this terminal, debris should be prevented from entering the pipeline system. During welding, attention should be paid to cooling the intake pipe to avoid damage to the terminal socket caused by high welding temperatures. Due to the oxidation of copper during welding and the generation of harmful substances, gas shielded welding should be used during welding.
After the installation is completed, oil free air or nitrogen should be used for blowing to eliminate the exhaust gas in the pipe.
When using this terminal, open flames are strictly prohibited. After encountering oxygen, an open flame can accelerate combustion or explode, posing a fire hazard.